Clean the workshop and filter tube regularly and use clear water and fine cloth when wiping. If you are testing perspiration fastness standard, you need to clean the external filter tube and inner box every 200 hours to remove the inorganic salt that is attached to the surface.
It is recommended to change the air filter (paper net) every 3 months, depending on the filth of the filter. Clean air filter (metal mesh) - check the air filter for dust, sundries. If you are in a dusty environment, you should clean it every two weeks, using a vacuum cleaner, or you can clean it with compressed air or tap water.
The irradiance calibration (information prompt M31) is conducted for every 500 hours of the equipment, and the calibration tool is the CR20 irradiance calibration meter;
The blackboard (black mark) thermometer is calibrated every six months. It can be done by customers themselves or by professional engineers.
Case temperature and humidity sensor probe needs to be replaced periodically, and the recommended period of use is one year.When the ISO 105 B02 or GB/T 8427 is executed, the IR filter needs to be changed in accordance with the standard for 500 hours.
Random calibration measurement tools (such as: CR20 irradiance calibration meter) need to be on a regular basis (every 12 months) use mail in Shanghai luo technology development co., LTD. (Q - LAB company Shanghai technology center) for tracing measurement calibration again.
Standard xenon lamp is recommended for 1500 hours of service life (information hint M14). Extra care should be taken when replacing the lamp tube and taking care of the light. When replacing the lamp tube, check and clean the lamp tube to trigger the electrode.
Equipment each running 2000 hours, there will be (message M15) alarm prompt, then please to a comprehensive cleaning and maintenance of equipment, include: machine internal water tank cleaning, replacement of air filters.When the machine is out of operation for a long time, it will be turned on for a short time.