Atlas Xenn Lamp Aging Test Chamber Maintenance

Clean the workshop and filter tube regularly and use clear water and fine cloth when wiping. If you are testing perspiration fastness standard, you need to clean the external filter tube and inner box every 200 hours to remove the inorganic salt that is attached to the surface.

It is recommended to change the air filter (paper net) every 3 months, depending on the filth of the filter. Clean air filter (metal mesh) - check the air filter for dust, sundries. If you are in a dusty environment, you should clean it every two weeks, using a vacuum cleaner, or you can clean it with compressed air or tap water.
The irradiance calibration (information prompt M31) is conducted for every 500 hours of the equipment, and the calibration tool is the CR20 irradiance calibration meter;
The blackboard (black mark) thermometer is calibrated every six months. It can be done by customers themselves or by professional engineers.
Case temperature and humidity sensor probe needs to be replaced periodically, and the recommended period of use is one year.When the ISO 105 B02 or GB/T 8427 is executed, the IR filter needs to be changed in accordance with the standard for 500 hours.
Random calibration measurement tools (such as: CR20 irradiance calibration meter) need to be on a regular basis (every 12 months) use mail in Shanghai luo technology development co., LTD. (Q - LAB company Shanghai technology center) for tracing measurement calibration again.
Standard xenon lamp is recommended for 1500 hours of service life (information hint M14). Extra care should be taken when replacing the lamp tube and taking care of the light. When replacing the lamp tube, check and clean the lamp tube to trigger the electrode.
Equipment each running 2000 hours, there will be (message M15) alarm prompt, then please to a comprehensive cleaning and maintenance of equipment, include: machine internal water tank cleaning, replacement of air filters.When the machine is out of operation for a long time, it will be turned on for a short time.

评论

此博客中的热门博文

How To Process The Elmendorf Tear Strength Test

The Elmendorf tear tester has been developed specially for the paper industry and is outstanding for simplicity of operation and high measurement accuracy. The exchangeable pendulum is equipped with a pneumatic sample clamp. The initiation and return of the pendulum, as well as sample cutting is completely automatic. The device is operated via an integrated touch screen, where the test method can be selected, specific settings can be adjusted, and, after the measurement, results and statistics can be  displayed.Each pendulum is supplied with the appropriate control weight so that, with the help of the service program, it can easily be checked for functionality and balance. The Elmendorf tear test Procedure:     Cotton fabric was taken.     Warp and weft direction of the fabric was identified.     Two strips were cut from the fabric in both directions having dimensions of 100 x 63.5 mm.     The sample was then fixed in the jaws of the ...

Comparative analysis of water-cooled and air-cooled xenon lamp aging tester

Xenon lamp aging test chamber is the core component of using the box full sunlight spectrum xenon lamp tube simulation, under different environmental conditions to simulate the material aging law, through the experimental results to evaluate weather resistance of materials and products. This paper focuses on the comparative analysis of the advantages and characteristics of the water-cooled and air-cooled xenon lamp aging testing machine. Air-cooled xenon lamp aging test machine : A circular wind is used to cool the heat-trapping xenon lamp tube. Air-cooled xenon lamp tubes used in air-cooled xenon lamp aging test box, generally a typhoon cold xenon lamp aging test chamber is equipped with three air-cooled xenon lamp tubes, and table 1 air-cooled xenon lamp weather-resistant chamber equipped with xenon lamp tubes. The power of the air-cooled xenon lamp tube is generally 1.8kw, and the length of the lamp tube is 340mm compared with that of the water-cooled xenon lamp. The diameter of...

Correct Use Of Color Assessment Cabinet

Observation Angle: The following points are all the facilities required for the Color Assessment Cabinet , and the viewing Angle will directly affect the evaluation results. According to international standards, there are only two observable angles that can be used. They are: A, zero light source and 45 degrees observation (0 °, 45 °), the light from zero (vertical) incident Angle on the sample, sample watchers from 45 degrees. B, light source and 45 ° inclined channel 45 degrees of light source, zero observation (0 °, 45 °), the layout must use specific 45 degrees of inclined (45 ° Fixed Angle table) making from 45 degrees of light source on the samples, sample watchers from zero (vertical). The location of the 45 ° light samples: Regardless of any of the above observation methods, it is important to note that the test sample is placed in the middle of the light box as far as possible to reduce the influence of the external light source. In addition, the most important thing...