跳至主要内容

Masks Flame Retardant Tester

 introduction:

The flame-retardant tester of medical mask is used to test the flame-retardant performance of the mask when it is accidentally contacted by flame, which is directly related to the safety of medical staff. The flame retardancy of mask is one of the important physical indexes to measure the quality of medical mask.
 
Application:
Mask Flame Retardant Tester is a special test instrument for flame retardant performance of medical masks,  to test the flammability performance of medical mask after contacting with flame at a certain linear speed.
 
Standards:
GB 19083, YY 0469, GB 2626, EN 149, IS 9473, EN 14325
 
FEATURES:
PLC + touch screen control system makes the test more convenient, operated and visualized;
The mask clamp is a metal human head mold, which can fully simulate the actual use state of the mask.
Burner height is adjustable.
Automatic timing and positioning of burner.
Combustion time and flame retardant time are recorded automatically and displayed digitally
Equipped with flame temperature measuring probe and flame temperature display.
Automatic control system, automatic ignition
The burner is equipped with LPG interface
Automatic timing, automatic storage of test data
 
parameter:
Test mode: Automatic
Burner timing: 0~9999.9s
Automatic linear velocity of head mould: (60±5)mm/s
Distance between the top of burner and the lowest point of mask: (20±2)mm
Flame height: 40±4mm
Temperature of flame at a height of (20 ± 2) mm from the top of burner: (800 ±50)℃
Temperature sensor: Thermocouple detection
Burner height adjustment range: 30mm
Metal die wall thickness: 2 ~ 3mm, flame resistant and flame resistant
Weight: 65kgs
Power supply: 220V 50Hz
Appearance size: 810mm * 510mm * 870mm

评论

此博客中的热门博文

How To Process The Elmendorf Tear Strength Test

The Elmendorf tear tester has been developed specially for the paper industry and is outstanding for simplicity of operation and high measurement accuracy. The exchangeable pendulum is equipped with a pneumatic sample clamp. The initiation and return of the pendulum, as well as sample cutting is completely automatic. The device is operated via an integrated touch screen, where the test method can be selected, specific settings can be adjusted, and, after the measurement, results and statistics can be  displayed.Each pendulum is supplied with the appropriate control weight so that, with the help of the service program, it can easily be checked for functionality and balance. The Elmendorf tear test Procedure:     Cotton fabric was taken.     Warp and weft direction of the fabric was identified.     Two strips were cut from the fabric in both directions having dimensions of 100 x 63.5 mm.     The sample was then fixed in the jaws of the ...

Comparative analysis of water-cooled and air-cooled xenon lamp aging tester

Xenon lamp aging test chamber is the core component of using the box full sunlight spectrum xenon lamp tube simulation, under different environmental conditions to simulate the material aging law, through the experimental results to evaluate weather resistance of materials and products. This paper focuses on the comparative analysis of the advantages and characteristics of the water-cooled and air-cooled xenon lamp aging testing machine. Air-cooled xenon lamp aging test machine : A circular wind is used to cool the heat-trapping xenon lamp tube. Air-cooled xenon lamp tubes used in air-cooled xenon lamp aging test box, generally a typhoon cold xenon lamp aging test chamber is equipped with three air-cooled xenon lamp tubes, and table 1 air-cooled xenon lamp weather-resistant chamber equipped with xenon lamp tubes. The power of the air-cooled xenon lamp tube is generally 1.8kw, and the length of the lamp tube is 340mm compared with that of the water-cooled xenon lamp. The diameter of...

3 main functional values of the Burning Manikin System

3 main functional values of the Burning Manikin System 1) Research on the heat transfer mechanism of individual soldier's flame retardant equipment By simulating the battlefield fire conditions of different intensities, the changes of human skin surface heat flow and skin temperature by burning dummies, a good human body-clothing-fire environment conditions are created, and the individual soldiers are flame retardant The study of equipment heat transfer mechanism provides an important research method. 2) Research and screening of flame retardant fabrics for individual soldier equipment Quantitatively test the flame retardant properties of different individual soldier equipment fabrics through the burning dummy system, guide the research on flame retardant fabrics, and screen out flame retardant fabrics with excellent performance. 3) Structure and style design of individual soldier's flame retardant protective equipment The structure and style of individual soldier's flame r...