Synthetic Blood Penetration Tester for Medical Face Mask

 Features of Synthetic Blood Penetration Tester:

1. Mask fixture is metal human head mold, which can fully simulate the actual use state of the mask.
2. Metal human head mold can be rotated at a certain angle, which is convenient for sample clamping.
3. Specially designed injection control device, a certain volume of synthetic blood can be sprayed in the controlled time.
4. Instrument uses a gas source that can provide (700±25) kPa stable air pressure.
5. It can simulate the liquid injection velocity corresponding to the human blood pressure 10.6 kPa、16.0kPa 、21.3kPa (i.e .80 mmHg、120mmHg、160mmHg) for the test.
6. A fixed target plate is provided to block the high pressure edge part of the ejected liquid flow, only the steady flow part is sprayed on the area sample, which increases the accuracy and repeatability of the test.
7. This blood penetration testing equipment uses imported special aluminum wire drawing panel, equipped with metal keys, sensitive operation, not easy to damage.
8. 7 inch color touch screen display test parameters, menu-style operation mode, the degree of convenience comparable to the smartphone.
9. Core control components using Italian law company 32-bit multifunctional motherboard.

Video Description:

Technical Specification:
Spraying distance: 300mm-305mm adjustableNozzle length: 12.7mm
Space between front and rear plates: 0.6cmTarget plate bore diameter: 0.5cm
Control panel: Touch-toneTotal weight: 35Kg
Nozzle size: 0.84mmPower supply: 220V 50HZ
Spraying speed: 450cm/s, 550cm/s, 635cm/sAppearance size: 1000mm*500mm* 650 mm
Spraying pressure: Standard/10.6kPa, 16kPa,21.3kPa/adjustable 

Applicable Standards:
  • GB 19083-2010 Technical requirements of medical face masks - Anti-penetration property of 5.5 synthetic blood.
  • YY/T 0691-2008 Clothing for protection against infectious agents-Medical face masks - Test method for resistance against penetration by synthetic blood (fixed volume, horizontally projected)  
  • YY 0469-2011 Technical requirements of medical face masks– Blood penetration test equipment.
  • ISO 22609:2004 Clothing for protection against infectious agents - Medical face masks - Test method for resistance against penetration by synthetic blood (fixed volume, horizontally projected)  
  • ASTM F1862-07Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Know Velocity)

Applications of Synthetic Blood Penetration Tester:
Synthetic blood penetration tester for medical face mask is used for anti-synthetic blood penetration of medical surgical masks, and mainly used in the field of medical protection products. The blood penetration test equipment can also be used for other textile coating material determination.





评论

此博客中的热门博文

How To Process The Elmendorf Tear Strength Test

The Elmendorf tear tester has been developed specially for the paper industry and is outstanding for simplicity of operation and high measurement accuracy. The exchangeable pendulum is equipped with a pneumatic sample clamp. The initiation and return of the pendulum, as well as sample cutting is completely automatic. The device is operated via an integrated touch screen, where the test method can be selected, specific settings can be adjusted, and, after the measurement, results and statistics can be  displayed.Each pendulum is supplied with the appropriate control weight so that, with the help of the service program, it can easily be checked for functionality and balance. The Elmendorf tear test Procedure:     Cotton fabric was taken.     Warp and weft direction of the fabric was identified.     Two strips were cut from the fabric in both directions having dimensions of 100 x 63.5 mm.     The sample was then fixed in the jaws of the ...

Correct Use Of Color Assessment Cabinet

Observation Angle: The following points are all the facilities required for the Color Assessment Cabinet , and the viewing Angle will directly affect the evaluation results. According to international standards, there are only two observable angles that can be used. They are: A, zero light source and 45 degrees observation (0 °, 45 °), the light from zero (vertical) incident Angle on the sample, sample watchers from 45 degrees. B, light source and 45 ° inclined channel 45 degrees of light source, zero observation (0 °, 45 °), the layout must use specific 45 degrees of inclined (45 ° Fixed Angle table) making from 45 degrees of light source on the samples, sample watchers from zero (vertical). The location of the 45 ° light samples: Regardless of any of the above observation methods, it is important to note that the test sample is placed in the middle of the light box as far as possible to reduce the influence of the external light source. In addition, the most important thing...

The Scratch Tester Working Principle From Qinsun LAB

The scratch test machine is used to evaluate the shear, scratch, planing, scraping and engraving performance of the material. Its applications include rigid organic materials, plastics and glass, and so on. The test parameters, such as the speed of scratches, the length of scratches, the distance between scratches and the number of lines, are the test equipment with strong applicability and convenience. scratch tester Working principle: Coating has three aspects: the action of protection, decoration, and hide the defects of products, and the three main hardness has a lot to do with paint, paint film hardness is said film one of the important properties of mechanical strength, is an important indicator of measuring coating product quality, and measure the coating hardness is an important index of the scratch resistance of determination; Coating good product in the process of packaging, transport and use will draw by hard brush, not row brush the paint film is often leave scratc...